OPERATING MANUAL
BFS-30VT REFRIGERATION MODULE
AND DIGITAL TEMPERATURE CONTROLLER
TABLE OF CONTENTS
1. Summary of Operation
2. Unpacking and Checking Package Contents.
3. Mounting the Refrigeration Module to VT1000 Microtome
4. Connecting the Controller to a Water Supply or the
Closed Loop Water System
5. Operating Instructions
6. Troubleshooting Information
7. Specifications
8. Warranty and Return Information
App.1 PTU-3 Operating Instructions
App.2 ECV-1 Operating Instructions
IMPORTANT
MUST ESTABLISH WATER FLOW
BEFORE TURNING THE UNIT ON
DO NOT IMMERSE THE BATH IN LIQUID OR ICE
1
EB BFS-30VT
Seitenansicht 0
1 2 3 4 5 6 ... 11 12

Inhaltsverzeichnis

Seite 1 - IMPORTANT

OPERATING MANUALBFS-30VT REFRIGERATION MODULEAND DIGITAL TEMPERATURE CONTROLLERTABLE OF CONTENTS1. Summary of Operation2. Unpacking and Checking Packa

Seite 2

APPENDIX 1OPERATING INSTRUCTIONS FOR PHYSITEMP PUMP AND TANK UNIT, PTU-31. Unscrew cap. Fill reservoir with 5 gallons of distilled water.2. Connect ex

Seite 3

APPENDIX 2OPERATING INSTRUCTIONSECV-1 ELECTRICALLY OPERATED SOLENOID VALVEThe ECV-1 is supplied with an AC line cord and 12 VDC desktop power supply.1

Seite 4

PHYSITEMP INSTRUMENTS INC154 Huron AvenueClifton NJ 07013 USATel: 973-779-5577Fax: 973-779-5954OPERATING MANUALEB FRONTIS EB BFS-VTBFS-30VT Bath Refri

Seite 5

1. Summary of OperationThe BFS-30VT is designed to provide consistent accurate cooling control to the VT-1000specimen bath. The BFS-30VT will hold tem

Seite 6 - CONDITION

3. Mounting the Refrigeration Module to yourVT1000 Sectioning SystemRemove the black refrigeration bath from its packaging. Remove the cutting head fr

Seite 7

Closed Loop Water System - PTU-3 Installation InstructionsThe PTU-3 is supplied with an AC line cord for connection tothe auxiliary outlet on the cont

Seite 8

5. Operating instructionsEnsure that the POWER switch on the rear panel of the controller is in the OFF position.Check the VOLTAGE SELECTOR switch on

Seite 9

Setting the module operating temperature4.6 SETTING THE STAGE TEMPERATUREThe operating temperature may be setanywhere in the range from ambientdown to

Seite 10 - GENERAL INSTRUCTIONS

6. Troubleshooting1. Controller display fails to illuminate when power is turned on.Check power cord continuity from the AC outlet to the input socket

Seite 11 - OPERATING INSTRUCTIONS

7. SpecificationsController SpecificationsSize: 7½ inches W x 10 inches D x 4 ½ inches H Weight: 5 poundsPower Requirements: 110-120V

Seite 12

8.0 MAINTENANCE, WARRANTY AND SERVICE8.1 MAINTENANCEThe stage needs no maintenace at all. It may be cleaned as necessary with a softcloth, water or de